30,120 Ventec V+ Pro emergency ventilators are under an FDA Class I recall

Ventec Life Systems, which operates as React Health, is recalling 30,120 Ventec V+ Pro Emergency ventilators in the United States, and the FDA has placed the action in its most serious recall classification. The defect is a flow transducer cable connector inside the machine that can work loose from its circuit board, causing the ventilator to reboot and interrupt breathing support. The recall covers every serial number of the model and reaches customers nationwide, including Puerto Rico.

Tech news worth reading: Get Techpresso, the daily email with the most interesting tech news and insights, read by 400,000+ professionals from Google, Apple and OpenAI.

Get Techpresso →

We may earn a commission if you sign up.

The company started the recall on Sept. 17, and the FDA recorded it in an enforcement report dated Oct. 28.

The record in the FDA’s openFDA device enforcement feed lists the product as the V+ Pro Emergency ventilator, model PRT-01201-000, with package part numbers PRT-01173-000 and PRT-01222-000, under recall number Z-0132-2027 and event 99835. It gives the quantity as 30,120 units and the status as ongoing. Ventec’s address in the record is Bothell, Washington. The FDA’s classification step came on Oct. 20, about a month after the firm acted.

By the FDA’s own definition, its most serious recall classification is reserved for situations in which there is a reasonable probability that a device will cause serious adverse health consequences or death. The classification does not mean harm has occurred, and the agency’s early alert said the chance of a reboot during active ventilation is remote. It does mean the agency judged the consequence of a failure severe enough to place the recall in the top tier.

The flow transducer cable behind the reboots

According to the FDA’s early alert posted Sept. 23, the internal flow transducer cable connector can partly or fully come loose, cutting communication with the flow transducer board. The result can be reboots and repeated System Fault 10 alarms. A substantial impact, such as dropping the device, can trigger the disconnection.

A reboot during active ventilation would interrupt therapy and could cause serious harm or death, especially for patients who cannot breathe adequately unaided. The agency considers that sequence remote, and as of Aug. 12 React Health had reported no serious injuries or deaths.

Timing shows how the process runs. React Health sent its customer letter on Sept. 17, the FDA posted its early alert six days later, the agency’s center classified the recall on Oct. 20, and the weekly enforcement report carried it on Oct. 28. The record page for sibling models was still being updated on Nov. 5, so counts and wording may continue to shift as the firm reports progress on corrections.

Four sibling models inside event 99835

The V+ Pro Emergency is one of several machines sharing the same defect and the same event number. The V+Pro for facility use is a larger group at 39,568 units under Z-0129-2027. The V*Home ventilator accounts for 25,112 units under Z-0133-2027, and the VOCSN+Pro for 4,369 units under Z-0122-2027.

A Japan-market version, the VC+Pro, is covered by 606 units under Z-0125-2027. Several of these records also list distribution in Japan and China, while the V+ Pro Emergency record is limited to the United States and Puerto Rico. The 30,120 figure in this article refers only to the Emergency model and should not be added to the others without checking for overlap.

Adding the five records read for this article gives 99,775 units, which is 30,120 plus 39,568, 25,112, 4,369 and 606. That total is only an upper bound for the recall as a whole, since the records are separate model lines and the notices do not say how many units, if any, appear twice. The early alert ties the issue to VOCSN-series devices with serial numbers 112786 and higher, which is a different way of drawing the boundary from the enforcement record’s all-serial-numbers language.

FDA and Ventec instructions for clinicians and caregivers

The agency’s guidance is practical. Keep backup ventilation working and ready, do not stop a patient’s therapy unless an alternative is available and a clinician directs it, and call React Health at 844-698-6276 for servicing. Secure the device on a flat, stable surface, the roll stand or the carry bag, and after any drop or impact run the checkout procedure in manual LBL-00001 before putting the machine back on a patient.

The practical stakes fall hardest on home and facility patients who depend on a ventilator for much of the day, because a backup device is the only protection during a reboot. That is why the FDA’s advice leads with keeping alternative ventilation on hand and working, ahead of any call to the manufacturer.

Caregivers should stay trained, monitor patients with pulse oximetry and carbon dioxide monitoring where available, and ship units only in approved packaging. The enforcement record says the FDA-determined cause is still under investigation by the firm and the recall remains open, so whether a repair or replacement schedule exists for the 30,120 Emergency units is not stated in the reports reviewed here.